Series WDPM Wireless Differential Pressure Module for UHH
Specifications - Installation and Operating Instructions
Bulletin TE-WDPM
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: [email protected]
SPECIFICATIONS
Service: Non-corrosive dry gases.
Wetted Materials: Consult factory.
Accuracy: ±0.5% FS span @ 25°C (includes non linearity, hysteresis, and non
r
epeatability).
P
ressure Limits: See Table 1.
T
emperature Limits:
Compensated: 32 to 140°F (0 to 60°C);
Process/Ambient: 14 to 140°F (-10 to 60°C).
Thermal Effects: ±0.01% FS/°F (±0.02% FS/°C).
Battery Charging Limits: 32 to 113°F (0 to 45°C).
Power Requirements: 3.7 V YT562447 lithium ion battery, installed functionally,
user replaceable.
Wireless Distance: At least 50´ (15 m).
Weight: 2.5 oz (70.87 g).
Connections: Two barbed connections for use with 1/8˝ (3.18 mm) or 3/16˝ (4.76
mm) ID tubing.
Agency Approvals: CE (not while charging), RoHS, FCC compliant.
The Series WDPM Wireless Differential Pressure Module transmits
measurements to the Model UHH handheld to view pressure drop across filters,
static pressure in ducts, and velocity pressures from pitot tubes or air flow stations.
When using it with a pitot tube or air flow measuring station, the measurement can
b
e calculated to read directly in air velocity or volumetric air flow. The module
m
ounts in the holder on the back of the handheld, or in an optional mounting
b
racket.
Wireless modules can take measurements from at least 50 feet away from the
base unit. An LED flashes on the top of the module to let the customer know when
the module communicates with the handheld, as well as when the module’s battery
is low. The battery is rechargeable via the mini-USB connector on the top of the
module.
CHARGING BATTERY
When the charge of the battery is almost used up, the LED on the module of the
wireless probe will turn solid red. There is approximately 5 minutes of battery life
left at this point. If the module is not already paired to the base unit, it will not pair
while in low battery condition.
Step 1: Plug the mini-USB connector end of the cable into the top of the module.
Step 2: Plug the USB connector end of the cable into the port on the charger or PC
(LED on the charger and the module should both light up).
Step 3: The LED on the module will turn off when fully charged.
Step 4: Remove USB cable from the module and the charger or PC.
It is required before the initial use of the module to fully charge
the battery for 12 hours.
NOTICE
Lithium ion batteries are very volatile and can cause a fire if
punctured or severely damaged. Only use a Dwyer Instruments,
Inc. approved charging device in a well ventilated area away from any flammable
materials or gases. Do not incinerate the battery. Only charge between 32 to 113°F
(0 to 45°C).
WARNING
If desired, it can be operated with USB powered cables less
than 3 m in length when connected to the charger or PC.
NOTICE
2-21/64
[59.28]
3
9/64
[
15.59]
1
-1/64
[
25.80]
3
-17/64
[
82.77]
2-21/32
[67.31]
Model
WDPM-002
WDPM-005
WDPM-010
WDPM-020
WDPM-030
WDPM-100
WDPM-200
WDPM-400
Range
±2 in w.c.
±5 in w.c.
±10 in w.c.
±20 in w.c.
±30 in w.c.
±100 in w.c.
±200 in w.c.
±400 in w.c.
Maximum Pressure
10 psi
10 psi
10 psi
20 psi
20 psi
15 psi
45 psi
45 psi
Table 1: Maximum Pressure
This device complies with Industry Canada license-exempt RSS
standard(s). Operation is subject to the following two conditions:
(1) This device may not cause harmful interference. and (2) this device must accept
any interference received, including interference that may cause undesired
operation.
Cet appareil est conforme à des règlements d'Industrie Canada exempts de licence
standard RSS (s). Son fonctionnement est soumis aux deux conditions suivantes:
(1) Ce dispositif ne doit pas causer d'interférences nuisibles, et (2) cet appareil doit
accepter toute interférence reçue, y compris les interférences pouvant entraîner un
fonctionnement indésirable.
This Class B digital apparatus complies with Canadian ICES-003.
Cet appareil numériqué de la classe B est conformé à la norme NMB-003 du
Canada.
NOTICE
Vue de la page 0
1 2 3 4

Résumé du contenu

Page 1 - DWYER INSTRUMENTS, INC

Series WDPM Wireless Differential Pressure Module for UHHSpecifications - Installation and Operating InstructionsBulletin TE-WDPMDWYER INSTRUMENTS, IN

Page 2

PAIRING WIRELESS MODULES1. Turn on Model UHH Universal Handheld by pressing the button.2. Press the and buttons to scroll through the

Page 3

LOGGING MEASUREMENTSThe UHH series can log the data being measured either a single point at a time orby continuously logging at a user defined sample

Page 4

DWYER INSTRUMENTS, INC. Phone: 219/879-8000 www.dwyer-inst.comP.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwye